AV-2000B2 is a multi-functional Ventilator, adopting pneumatic, electronic control and microcomputer technology. Imported components are utilized to guarantee stablility and reliabiligyt. With color screen(TEF) to display various measuringand configuration parameters, the ventilator suitable for respiratory treatment, applicable for medical,surgical and gynecological department, pediatrics department, first aid, ICU etc.Air source cen be conneted.

Need Help?
Call Us +90 507 111 87 89
CONTACT ME ON WHATSAPP
  • Shipping modes: Express (air), economy air, sea (FCL/LCL), and road for regional moves.

  • Lead times: Stock items 2–5 business days; made-to-order 10–25 days (depends on model & quantity).

  • Incoterms: EXW, FCA, FOB, CFR/CIF, DAP, DDP (on request).

  • Carriers & lanes: Dedicated medical logistics partners with global coverage.

  • Tracking: End-to-end tracking with milestone emails; POD shared on delivery.

  • Documentation: Commercial invoice, packing list, HS codes, CE/ISO declarations, test reports (where applicable).

  • Customs & compliance: Integrated brokerage; pre-clearance support and country-specific labeling.

  • Packaging: Sterile, sealed inner packs; shock-/moisture-protected export cartons; ISPM-15 pallets.

  • Insurance: Cargo insurance available upon request.

  • Support: WhatsApp / Email for live status & quick changes.

Description

Parameter

Ventilator:

Display mode

High-definition 8.4” TFT color LCD screen display

Ventilation  mode

IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH

Ventilation parameters

Tidal volume :       50~1500 ml

Rate :              2~99 bpm

SIMV rate :         2~20 bpm

I:E :               2:1~1:8

Inspiratory trigger pressure: -1.0~2.0kPa

PEEP:              0~2.0 kPa

Pressure range :      1.0~6.0 kPa

SIGH :    1.5 times the inspiratory time

Inspiratory platform :  0~50%

Parameters for ventilation monitoring

tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure – time waveform, flow rate – time waveform, PEEP, inspiratory trigger pressure, Inspiratory platform

Monitoring of oxygen concentration

21%~100%

Safety alarm system : 

Oxygen Concentration alarm

upper limit setting range 50%~100%

low limit setting range   15%~50%

Airway pressure alarm

upper limit setting range   1.0~6.0 kPa

low limit setting range    0.4~2.0 kPa

Per-minute ventilation volume alarm

upper limit setting range 3.0~30L/min

low limit setting range  1.0~10L/min

Sustained high-pressure alarm

it will give alarm when stress have consistently

been higher than 2.5 kPa

Suffocation alarm

it will give sound and light alarm if there is no tidal volume input for 15 seconds

Power alarm

Power 

AC 220V±10%     50Hz±1Hz

 UPS, Storage battery

Video